3.6.6 UTILITY Menu

Utility Menu


F1: GPIB

-Selects the GPIB remote mode of operation. After selection the GPIB address can be set to any
value from 1 to 31 using the rotary knob. The value is kept in a nonvolatile memory and used at
power-on. The factory default address is 10. Setting the address to 31 puts the device in the off-
bus state (it will not respond to messages on the GPIB bus).

GPIB Menu

F2: RS232

-Selects the RS232 remote control mode. After selection, the baud rate can be selected as 1200,
2400, 9600, 19200, 38400, 57600 or 115K. Always the RS-232 uses 8 bit data, 1 stop bit and no
parity.

Note: If 115K baudrate speed is used, ensure that the RS232 cable is short and can support this
speed. Otherwise, there may be some instability and intermittent data transmission failure
between the interfacing computer and the instrument.


F3: INTEN

- Selects the intensity of the LCD display. Select a value using the rotary input knob. Valid
numeric values are from 1 to 31. The value is kept in the nonvolatile memory, after a 20 seconds