56

is generated after a programmable delay. This delay is set by either the :PULSE:DELAY or the
:PULSE:DOUBLE:DELAY command.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:PULSe:DOUBle[:STATe]<ws><Boolean>

Arguments:

Type:

Boolean

Examples:

:PULS:DOUB ON

:PULS:DOUB:STAT OFF

QUERY

Syntax:

[:SOURce]:PULSe:DOUBle[STATe]?

Examples:

:PULS:DOUB?

Response:

0 | 1

4.12.1.16 Double Pulse Delay

This command is used to set the delay of the second pulse, from the time of the trigger, in the double
pulse mode. It has exactly the same effect as the :PULSE:DELAY command, and is included in the
command set for compatibility purposes.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:PULSe:DOUBle:DELay<ws><delay>[units]

Arguments:

Type:

NRf

Units:

S (seconds), MS (milliseconds), US (microseconds), NS nanoseconds

Range:

0NS to 9.80000S

Rounding:

To current resolution

Examples:

:PULS:DOUB:DELay 150NS

QUERY

Syntax:

[:SOURce]:PULSe:DOUBle:DELay?

Examples:

:PULS:DOUB:DEL?

Response:

NRf

CONSIDERATIONS:

The allowed range of the delay will be determined by the values of the period, width, and
transition times.

4.12.1.17 Leading Edge Time

This command is used to set the value of the leading edge time. If the edge-tracking feature is ON,
changing the leading edge will cause the same change in the trailing edge.

COMMAND TYPE:

Setting or Query

SETTING

Syntax:

[:SOURce]:PULSe:TRANsition[:LEADing]<ws><lead time>[units]