102

Track Mode

Table 33- Track Mode Function Menu

Option

Setting

Instruction

Cursor
Mode

Track

In this mode, set track
cursor measure.

Cursor A

CH1
CH2
NONE

Set the input signal
channel that the Cursor A
will measure.

Cursor B

CH1
CH2
NONE

Set the input signal
channel that the Cursor B
will measure.

Cur A

Select this option, turn the
“Universal” knob to adjust
horizontal coordinate of
Cursor A.

Cur B

Select this option, turn the
“Universal” knob to adjust
horizontal coordinate of
Cursor B.


In this mode, the screen displays two cross cursors. The cross cursor sets

the position on the waveform automatically. You could adjust the cursor’s

horizontal position on the waveform by turning the “Universal” knob. The

oscilloscope displays the values on the top left of the screen.

To do track cursor measurement, follow these steps:

1. Press CURSOR button to enter the cursor measure function

menu.

2. Press the “Cursor Mode” option button to select “Track”.

3. Press the “Cursor A” option button to select the input signal

channel.