Self calibration – B&K Precision 2532B - Manual User Manual

Page 138

Advertising
background image

138

Figure 70 - Language Selection Screen



Self Calibration

Self Calibration procedure can optimize the signal path at the most

measurement precision. You can run this procedure at any time. If the

operating temperature changes by more than 5° C or the units run more

than thirty minutes, you should do the self calibration.

When you run the self Calibration, you should disconnect all the

probes and leads, and make sure the scope is not in Single-Shot mode.

Then press the “Utility” button to choose the “Do self cal” to show the

self Calibration Menu, and execute self calibration program according

to the prompt on the screen.