Figure 34 - Video Trigger Screen

Table 18 - Functional Menu of Video Trigger 2

Option

Setting

Instruction

Type

Video

When you select the video
type, put the couple set to
the AC, then you could
trigger the NTSC,PAL and
SECAM video signal.

Standard

NTSC
Pal/Secam

Select the video standard for sync
and line number count.

Mode

Auto

Use this mode to let the
acquisition free-run in the absence
of a valid trigger.

Normal

Use this mode when you want to
see only valid triggered
waveforms; when you use this
mode, the oscilloscope does not
display a waveform until after the
first trigger.

Single

When you want the oscilloscope
to acquire a single waveform,
press the “SINGLE ”button.

Set up

Enter the “Trigger setup menu”.