“GND” Coupling: Use GND coupling to display a zero-volt

waveform. Internally, the channel input is connected to a zero-volt

reference level.

Fine Resolution: The vertical scale readout displays the actual

Volts/Div setting while in the fine resolution setting. Changing the

setting to coarse does not change the vertical scale until the

VOLTS/DIV control is adjusted.

Setting Up Channels

Each channel has its own separate Menu. The items are set up

separately according to each channel.

1. Setup Channel Coupling

Note

● The oscilloscope vertical response rolls off slowly above its

bandwidth. Therefore, the FFT spectrum can show valid frequency

information higher than the oscilloscope bandwidth. However, the

magnitude information near or above the bandwidth will not be

accurate.

● If the channel is set to DC coupling, you can quickly measure the

DC component of the signal by simply noting its distance from the

ground symbol.

● If the channel is set to AC coupling, the DC component of the

signal is blocked allowing you to use greater sensitivity to display

the AC component of the signal.