Trigger holdoff – B&K Precision 2532B - Manual User Manual

Page 82

Advertising
background image

82

The TRIGGER LEVEL knob controls where on the edge the
trigger point occurs

.

Figure 38 - Rise and Fall Edge









Trigger Holdoff

You can use the Trigger Holdoff function to produce a stable display of

complex waveforms. Holdoff is time between when the oscilloscope

detects one trigger and when it is ready to detect another. The oscilloscope

Note:

● Press the SINGLE button when you want the oscilloscope to

acquire a single waveform.

● Trigger coupling affects only the signal passed to the trigger

system. It does not affect the bandwidth or coupling of the

signal displayed on the screen.

● Normal Polarity Sync triggers always occur on negative-going

horizontal sync pulses. If the video waveform has positive-going

horizontal sync pulses, use the Inverted Polarity selection.

Advertising
This manual is related to the following products: