Page 98

Advertising
background image

98

the horizontal scale and position.

Turn the “time/div” knob to adjust the sampling rate

The following functions are forbidden at XY display

form:

● Benchmark waveform and wave mathematics

Cursor

● Auto (resets display format to YT)

Trigger Control

Horizontal Position Knob

Vector Display Type

Scan Display











Note: When the display type is set to Vector, the oscilloscope

connects the sample points by using digital interpolation.

Interpolation mode contains linear interpolation and sin(x)/x

interpolation.

Sinx/x interpolation is suitable for reconstructing

curved or irregular wave shapes, especially when the sample

rate is 3 to 5 times the system bandwidth

or at faster time base

settings.