6.

If you are completing the steam sterilization for the first time, plug
the water distiller into a dedicated wall outlet.

Steam Sterilization Operation Switches Diagram

7.

Refer to Steam Sterilization Operation Switches Diagram
above. Turn only the power and water switches ON. Leave the fan
switch OFF. As the water distiller continues to heat up, it will
sterilize the water distiller with steam by allowing pure steam to
travel out of the cooling coil. Water and steam will drip from the
sterilization tube. Caution: The Water Distiller is extremely hot
during sterilization.

8.

Leave the water distiller on for 1 hour.

9.

After sterilization cycle is complete. Turn the fan switch ON and
allow any water in the cooling coil to drain. Leave the water distiller
running with the fan on for about 15 minutes.

10.

The water distiller will be very hot so turn all switches OFF (power,
water, fan) and let the water distiller cool down for approximately 1
hour.

11.

Remove the sterilization drip tube by pulling the plastic tube off the
stainless steel charcoal filter inlet tube (keep the sterilization tube
for future sterilizations).

12.

Replace the charcoal in the charcoal filter canister. See Replacing
Charcoal in Charcoal Filter Canister
in this Maintenance and
Cleaning Section.

13.

The water distiller head is sterilized and ready for operation.