6.3.3.11 Function set to SYSTEM ERROR working mode

Figure 65 Timer mode—Relay behavior

1

OFF Duration

3

OFF Delay

2

Duration

4

Time (x-axis)

Table 37 Color/line code for

Figure 65

Relay contact (DUTY CYCLE=DIRECT)
Relay contact (DUTY CYCLE=REVERSE)

SYSTEM ERROR

WARNING LIST

Default value: DISABLED
ENABLED: Monitors the internal warning bits of every probe.
DISABLED: Monitoring is disabled.

ERROR LIST

Default value: DISABLED
ENABLED: Monitors the internal error bits of every probe.
DISABLED: Monitoring is disabled.

SENSOR MISSING

Default value: DISABLED
Monitors the connection of every connected probe.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

ON DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay for the relay to turn on.

OFF DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay for the relay to turn off.