Advanced operations – Hach-Lange SC 1000 User Manual

Page 85

Advertising
background image

83

Advanced Operations

ANALOG CURRENT INPUT
The current input card connects devices with a current input interface to the sc1000
controller. Every current input channel can be configured separately, unit and parameter
are displayed in the measured value display. It is mandatory to have an corresponding
open jumper on the current input card to connect a device.

DIGITAL CURRENT INPUT
To differ two digital states the corresponding jumper on the internal current input card has
to be closed respectively the corresponding bridge has to be set at the external current
input card. The different states are recognized by closing or opening a contact between
the corresponding screwing terminals.

There is the possibility to adjust the input current measuring with an offset and a
correction factor to enhance the accuracy. By default these two parameters are set to “0”
(offset) and “1” (correction factor). When a channel is used as a digital input the display
will show the values “HIGH” or “LOW”.

sc1000 SETUP
CURRENT INPUTS
mA INPUT INT/EXT

Select INPUT card 1,2,3 or 4

EDIT NAME

Default value: Serial number of device as text
Enter text for example for the location of the current source.

DEVICE NAME

Default value: No text
Sets the device name.

PARAMETER NAME

Default value: No text
Sets the parameter name.

SET PARAMETER

Default value: “ChanX” (X=Channel number of the input current module)
Sets parameter for the calculated output value.

DATA VIEW

Default value: OUTPUT VALUE
Sets the value which is displayed as the measured value in the display module and logged into
the data logger.

INPUT CURRENT

Displays the real measured input current.

OUTPUT VALUE

Displays the calculated output value after scaling the output value with the SET LOW VALUE
and SET HIGH VALUE menu settings.

UNIT

Default value: No text
Sets the unit for the calculated output value.

SET FUNCTION

Default value: ANALOG

ANALOG

Input channel is used as analog input.

DIGITAL

Input channel is used as digital input.

SET FILTER

Default value: 10 seconds
Sets a period of time for recording measured input currents.
The input current is the result of an average value, calculated from the last measured input
currents recorded over a defined period (which is set in this menu).

LOGIC

Default value: DIRECT
Sets the relation between input state and output level.
The menu item is displayed if SET FUNCTION is set to DIGITAL.

DIRECT

If the input contact is closed the output level is LOW respectively if the input contact is open the
output level is HIGH.

REVERSE

If the input contact is closed the output level is HIGH respectively if the input contact is open the
output level is LOW.

SCALE 0 mA/4 mA

Default value: 0–20 mA
Sets the input current range either to 0–20 mA or 4–20 mA.