Troubleshooting – Hach-Lange SC 1000 User Manual

Page 129

Advertising
background image

127

Troubleshooting

6. Cross-check the small input current.

7. Repeat the steps 1–6 until the input precision reaches a desired value.

TEST/MAINT
CURRENT INPUTS
mA INPUT INT/EXT

FUNCTION TEST

INPUT 1-4

INPUT CURRENT Displays the input current according to the current set to the corresponding channel.

SET FACTOR

Default value: 1
Adjusts the displayed input current with an factor.

SET OFFSET

Default value: 0
Adjusts the displayed input current with an offset.

OUTPUT MODE

Default value: HOLD
Decides how other members react if they try to read an input current value while the input
current card is in the function test. As the value of the input current in the function test is not
based on any measurements other members reading this value probably have to be informed
about this special situation. There are three settings: Hold, Active and Transfer.

HOLD

Another reading member does not use the current read value but uses the last value before the
output current card was in the function test mode.

ACTIVE

Another reading member uses the current value even if the output current card is in the function
test mode.

TRANSFER

The reading member uses an own substitute value for its own output value.