2 error and status register, Standard operations – Hach-Lange SC 1000 User Manual
Page 66
64
Standard Operations
7. Select a tag and press the
ENTER
button. The new tag is added to the Telegram list.
Select a tag and press the
UP
and
DOWN
button to move the position of the tag
8. Repeat the steps to add further devices and tags.
9. Press the
ENTER
button to store the Profibus configuration.
5.12.2 Error and status register
Note: ERROR and STATUS definitions are valid for all sc probes.
Figure 48 Profibus/Modbus configuration menu
—
Telegram list with new tag
Table 15 Telegram list—Column description
Column
Description
1
Profibus: Data position in the configured Profibus slave (in 2 byte words)
Modbus: Data position in the configured Modbus slave
This slave contains holding registers beginning at 40001.
Example: “0” means register 40001 or “11” means register 40012.
2
Tag name to identify the configured data.
3
Data type
float=floating point value
int=integer numbers
sel=integer value resulting from a enum (selection) list
4
Data status
r=data are read only
r/w=read/write