2 test the input card, Troubleshooting 8.5.2 test the input card – Hach-Lange SC 1000 User Manual

Page 128

Advertising
background image

126

Troubleshooting

8.5.2 Test the input card

In the TEST/MAINT menu, every input current channel can be checked. If necessary,
each input can be adjusted as well.

The input current can be checked by setting a defined current to the according channel
and then comparing it with the displayed value. If necessary, the displayed value can be
adjusted with an offset and a factor.

To set these parameters:

1. Set the SET OFFSET value to “0” and the SET FACTOR value to “1”.

2. Set the input current to a fairly small value (1 mA for example).

3. Adjust the SET OFFSET until the displayed current meets the set current.

4. Set the input current to a fairly big value (19 mA for example).

5. Adjust the SET FACTOR value until the displayed input current meets the set input

current.

TEST/MAINT
OUTPUT SETUP
mA OUTPUT INT/EXT

FUNCTION TEST

OUTPUT 1-4

CURRENT OUT

Select an output current and set at the appropriate output.

SET FACTOR

Default value: 1
Adjusts the output current with an factor changed with this value.

SET OFFSET

Default value: 0
Adjusts the output current with an offset changed with this value.

ALL

Default: 0 mA
Sets OUTPUT 1-4 to 0, 4, 10, 12 or 20 mA.

SET OUTMODE

Default value: HOLD
Decides how other members react if they try to read an output current value while the output
current card is in the function test. As the value of the output current in the function test is not
based on any calculations, other members reading this value probably have to be informed
about this special situation.

HOLD

Another reading member does not use the current read value, but uses the last value before the
output current card was in the function test mode.

ACTIVE

Another reading member uses the current value even when the output current card is in the
function test mode.

SET TRANSFER

The reading member uses an own substitute value for its own output value.

STATUS OUTPUT

Indicates the status for each output current channel and probe which is read from the output
current card.

SENSOR OK

The according output channel is working fine and the current output card is able to read the data
from the probe to set the output current.

SENSOR MISSING

The according output current channel cannot get data from the probe because the probe is not
responding anymore. In this case the output current has the value set in sc1000 SETUP>SET
TRANSFER or holds the according current value read the last time the probe was responding.

SENSOR FAIL

The output current card communicates with the according probe, but this probe has got an
internal failure and cannot provide reliable data. In this case the output current has the value set
in the menu SET TRANSFER in sc1000 SETUP or holds the according current value read the
last time the probe was responding.