107

Advanced Operations

6.3.4 Network Modules (Profibus, Modbus)

The sc1000 controller can be embedded as a slave into an existing fieldbus system. The
Network Modules menu displays all the settings needed, the menu content depends on
the communication gateway used, Profibus DP or Modbus.

Note: This menu is displayed only if a network card is installed in the sc1000 controller.

6.3.4.1 Profibus

SYSTEM ERROR

LOG INTERVAL

Default value: OFF
Sets the interval for logging the displayed value into the data logger.
Options: OFF, 5 minutes, 10 minutes, 15 minutes, 20 minutes, 30 minutes.

sc1000 SETUP
NETWORK MODULES
FIELDBUS

TELEGRAM

Builds an individual composition of data from different devices. These data structure allows to
transmit up to 24 measured values on one Profibus slave device.
For details refer to

section 5.12, page 61

.

PROFIBUS DP

ADDRESS

Default value: 0
Sets up the PROFIBUS address (1 to 128) for the slave.

DATA ORDER

Default value: NORMAL
Sets the sequence of bytes when transmitting floating point values.
Please note that this setting only affects the data of the configured slave.
A floating point value consists of 4 bytes.
SWAPPED: Swaps the first pair of bytes with the last pair.
NORMAL: The pairs are not swapped. This mode fits to all known Profibus master systems.
Note: A wrong setting in this menu can lead to slight deviations of floating point values shifted by
one register.

SIMULATION

Simulates two floating point values and error/status to substitute a real instrument.
The tag order is:
1. Tag: ERROR
2. Tag: STATUS
3./4. Tag: First floating point value counting to MAXIMUM value respectively MINIMUM value.
5./6. Tag: Second floating point being the difference between the first floating point value and the
value set in the MAXIMUM. menu
The first floating point runs through a ramp between the limits set in the menus MAXIMUM and
MINIMUM.

Figure 66

shows the simulation mode.

SIMULATION

Default value: NO
Turns the simulation on or off.
YES: Starts a simulation
NO: Stops a simulation.

PERIOD

Default value: 10 minutes
Sets the time the first floating point value needs to run through the whole range between
MINIMUM and MAXIMUM.

MAXIMUM

Default value: 100
Sets the upper limit for the first floating point value.

MINIMUM

Default value: 50
Sets the lower limit for the first floating point value.