3.7.1 sc1000 network connections

To attach a network connector:

1. Strip the insulation from the communication cable (

Figure 30

).

2. Feed the cable through the union nut, rubber seal, and connector housing

(

Figure 32

).

3. Connect the cable to the network connector circuit board as shown in

Table 11

.

Network Connector Assembly

4. Place the circuit board with the cable connected in the bottom part of the metal

frame.

5. Tighten the cable connector.

6. Place the top of the metal frame on the bottom and press together.

7. Feed the frame into the sc1000 connector. The frame will only fit in one position. If

necessary, rotate the frame.

8. Attach the circuit board and the frame to the front with the two self-tapping screws

provided.

Figure 29 Plugging the network connector to the network interface