Advanced Operations

The output value (OV) is a function of the input current (IC).

The output value is determined according to formula

(3)

:

where:
OV=Output value
IC=Input current
LV=Low value
HV=High value

Figure 54

shows the output value depending on the input current, the set low value and

the set high value with an input range of 4

20 mA.

Figure 53 Output value with an input range of 0

20 mA

1

Output value (concentration) (x-axis)

5

0 mA

2

OV=f(IC)

6

0 mA

3

Input current (IC) (y-axis)

7

0 mA

4

20 mA

8

0 mA

(3)

OV

=

f(IC)

=

IC

×

HV – LV

+

LV

20 mA