Advanced Operations

Relation between input current and calculated concentration

Figure 53

shows the output value depending on the input current, the set low value and

the set high value with an input range of 0

20 mA.

SET HIGH VALUE

Default value: 20
Sets the value of the output value when the input current is 20 mA.

SET LOW VALUE

Default value: 0
Sets the value of the output value when the input current is 0 mA (0–20 mA scale) or 4 mA
(4–20 mA scale).

ON ERROR MODE

Default value: OFF
An error is reported when input current is out of range (this can be 0–20 mA or 4–20 mA).
When set to “OFF” no error will be reported even if the input current is out of range.

0 mA

Substitute value is 0 mA in case of error.

4 mA

Substitute value is 4 mA in case of error.

20 mA

Substitute value is 20 mA in case of error.

OFF

No substitute value is used to replace the measured value in case of an error.

CONCENTRATION

Displays the calculated concentration depending on the input current and the scaling which is set
in the SET LOW VALUE and SET HIGH VALUE menu.

LOG INTERVAL

Default value: 10 minutes
Sets the interval for logging the displayed value into the data logger.
Options: OFF, 5 minutes, 10 minutes, 15 minutes, 20 minutes, 30 minutes

VERSION

Displays the software version number

LOCATION

Displays the current location

sc1000 SETUP
CURRENT INPUTS