Re-run or recover incomplete batch -61 – Luminex 100 IS Version 2.3 User Manual

Page 103

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 61

Figure 5-31. Processing Multiple Plates

After the first microtiter plate in a multiple plate batch has been
acquired, the system pauses and prompts you to insert the next plate.

‹

To process multiple plates during a batch or multi-batch:

1. Create the batch or multi-batch. See “Create a New Batch” on

page 5-53 or “Create a Multi-Batch” on page 5-58.

2. Insert the first plate.

3. On the Run Batch tab, click Start Plate to begin processing the

batch. When the initial plate is through, the system pauses and
displays the Insert Next Plate message in red underneath the
sheath pressure gauge.

4. Click Eject, then remove the acquired plate and load the next

plate for processing.

5. Click Resume. The system resumes the acquisition process.

Re-Run or

Recover

Incomplete Batch

An incomplete batch can be caused by situations such as a power
failure, software failure, marking a batch for deletion, or clicking
Cancel All. Use this procedure to re-run or recover an incomplete
batch.