Standards tab -24, Standards tab – Luminex 100 IS Version 2.3 User Manual

Page 66

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 24

PN 89-00002-00-072 Rev. C

Standards Tab

The Standards tab lists all the batch tests with system comments
specific to each test. The detailed test information that is displayed
on the three tabs of the Analysis window is determined by the test
selected under the Standards tab.

The Standards tab displays the quantitative batch’s standard curve. It
displays qualitative batches as a graph plotting the assay standard.
Above the graphical information, the tab displays the formula that it
used to calculate batch data. Below the graphical information, the tab
displays the Standards and Controls values for the selected test.

Table 5-5. Error Categories

System and Assay Errors

Sample Errors

Instrument Not Calibrated

Failed Verification (system lists
each failed control)

Failed Control (verifier failed)

Temperature Divergence from
Calibration Temperature

Failed Curve Fit

Analysis Error

APD Temp Range Exceeded

XYP Temp Unstable

Low Voltage

High Sheath Pressure

Low Sheath Pressure

Command Timeout

Low Laser Power

Cannot Calculate Inverse
Function

Failed Std in Batch

Insufficient Bead Count

Sample High/Low

Analyzer Error

High Sheath Pressure

Low Sheath Pressure

Sample Timeout

Sample Empty

Cannot Calculate Inverse
Function

Different Qualitative Results