Fluid leaks, Fluid leaks -4 – Luminex 100 IS Version 2.3 User Manual

Page 158

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

7 - 4

PN 89-00002-00-072 Rev. C

Fluid Leaks

Fluid leaks can result in poor pressurization and failed sample
acquisition.

Problem internal to the
instrument.

Determine if the problem is with
the analyzer or the SD or bottles
by disconnecting the sheath line
from the analyzer and running a
prime. Check the air pressure on
the Diagnostic tab. If air
pressure builds, the problem is
with the SD or sheath fluid
bottle.

Pressure too High

The sheath bottle is overfilled.

Ensure that the sheath bottle is
not filled above the fill line.

Regulator not adjusted properly.

If using bottles, open the center
door on the Luminex analyzer.
Use a screwdriver to adjust the
regulator to fit in the center of
the green region on the Run
Batch tab.

Symptom

Possible problem

Solution

Pressure too low

The sample probe is clogged.

Clean the sample probe. See
page 6-4.

The syringe seal leaks.

Replace the syringe seal. See
page 6-8.

The syringe valve leaks

Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime
command. If leaks continue, call
Technical Support.

Large amount of fluid pooled
around instrument.

Fittings or fluid lines are
damaged.

Call Technical Support.

Fluid dripping from the
sample probe.

The sample probe is clogged.

Clean the sample probe. See
page 6-4.

The sample three-way valve is
faulty.

Contact Technical Support.

Fluid is leaking from the front
of the analyzer.

The syringe seal leaks.

Replace the syringe seal. See
page 6-8.

Advertising