Status bar, Status bar -17 – Luminex 100 IS Version 2.3 User Manual

Page 59

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 17

collected events. Post acquisition does not display decaying dot plot;
it’s only a real-time function.

Status Bar

The Status Bar displays information about the Command State,
Device Status, Device Activity, Laser Status, Total Events per
Second, and Region Events per Second. Color coding indicates the
urgency of each item’s status. Device Activity uses no color coding.

Figure 5-15. System Status Bar

Table 5-3 describes the types of status bar messages in relation to the
message color coding.

Table 5-3. Status Bar Color Coding

Category

Color

Indicates

Command
State

Indicates communication status of the Luminex analyzer or
operations being processed

Green

Idle or processing

Yellow

Connecting, pausing, or paused

Red

Disconnected or locked out

Device Status

Indicates the current process of or warning about the
Luminex analyzer

Green

Running or standby

Yellow

Busy, pressurizing, sheath is empty, or warming
up

Red

Not ready or disconnected

Device
Activity

Indicates the activity that the Luminex analyzer is
performing.

Note:

The device activity has no color code.