12. Connect the color-coded tubing connectors to the Luminex 100

analyzer. Blue = sheath, green = air, orange = waste.

13. Fill the sheath bottle with sheath fluid to just below the air intake

fitting. Place the sheath and waste containers to the left side of
the analyzer. Connect the blue connector to the sheath fitting
(bottom fitting) on the sheath bottle. Connect the green
connector to the air fitting (top fitting) on the sheath bottle.
Connect the orange connector to the waste fitting on the waste
bottle.

14. Plug the Luminex XYP instrument, Luminex 100 analyzer, PC,

and monitor power cords into approved outlets. We strongly
recommend using an uninterruptible power supply to protect the
system from power variations. Refer to Recommended
Additional Equipment (page 3-7) for addi
tional information.

Install the Luminex XYP
Instrument Sample Probe

‹

To install the Luminex XYP instrument sample probe :

1. Verify that the Luminex 100 analyzer and Luminex XYP

instrument power switches are off. Unplug them from the
electrical outlet.

2. Remove the light housing directly above the sample arm by

grasping and firmly pulling out. It remains attached by a wire
harness. Place it on top of the analyzer. See Figure B-6.

Figure B-6 Light Housing Removed

Note:

The sheath and waste

containers must be placed on the
same surface as the Luminex XYP
instrument. Do not place the
containers on top of the Luminex
100 analyzer.

1. Light Housing

2. Cheminert fitting with cap

1

2