Luminex 100 is 2.3 software installation – Luminex 100 IS Version 2.3 User Manual

Page 196

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

B - 14

PN 89-00002-00-072 Rev. C

‹

To install the Luminex XYP instrument heater block:

1. On the Maintenance tab, click Eject/Retract to eject the plate

holder.

2. Insert the Luminex XYP instrument heater block on the plate

holder. See Figure B-14.

Figure B-14 The Luminex XYP Instrument Heater Block

3. Click Eject/Retract to retract the plate holder.

4. Turn the heater on and set the temperature. Refer to Set Luminex

XYP Instrument Heater Temperature (page 5-45).

For new system installations, perform the preliminary commands
listed in the Prepare System for First Use (page B-26) before running
samples.

Luminex 100 IS 2.3

Software

Installation

New Luminex 100 IS 2.3 systems arrive with the Luminex 100 IS
2.3 software installed. Upgrading existing systems to Luminex 100
IS Version 2.3 requires that you upgrade your software through an
upgrade kit. The upgrade kits include necessary components, such as
software CDs, manuals, cables, and instructions.

Select the upgrade kit that applies to your current version or is called
out in your upgrade kit procedure. If your current software version is:

Luminex 100 Version 1.7 with Windows 98 you need Upgrade
Kit CN-U010-01. Follow the procedure on page B-15.

Luminex 100 Version 1.7 with Windows 2000 you need
Upgrade Kit CN-U010-06. Follow the procedure on page B-17.

Luminex 100 IS Version 2.1 or 2.2 you need Upgrade Kit
U010-07. Follow the instructions on page B-19.