After running samples, Routine tasks, Sheath and waste fluids – Luminex 100 IS Version 2.3 User Manual

Page 142: After running samples -2, Routine tasks -2, Sheath and waste fluids -2

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

6 - 2

PN 89-00002-00-072 Rev. C

5. Click Wash. In the confirmation dialog box, click on the drop-

down arrow located to the right of the Eject/Retract button to
choose the desired location. Click Eject/Retract. The Luminex
XYP instrument tray ejects. Place sheath fluid in the selected
well or reservoir on the plate. Click OK. Perform this step twice.

6. Check that the Luminex XYP instrument sample probe has been

vertically aligned for the plate used in the kit.

After Running Samples

‹

After running samples:

Refer to “Maintenance Procedures” on page 5-93 for detailed
Sanitize and Soak command operation.

1. Sanitize with a 10% to 20% household bleach solution.

2. Run two Wash cycles with distilled water.

3. Soak with distilled water. Wait until the soak completes.

4. If desired, turn off the Luminex 100 analyzer.

Routine Tasks

Sheath and Waste Fluids

Replace the sheath fluid and empty the waste container as required.
You must manually monitor waste container levels. The Run Batch
tab displays a warning when you need to refill (or replace) the sheath
fluid container. See Figure 6-1.

Figure 6-1 Refill Sheath Fluid Warning (Circled in Photo)