MAP Technology

The System

PN 89-00002-00-072 Rev. C

3 - 5

55°C (95°F to 131°F) when performing heated assays and using
the heater block

Luminex 100

Analyzer General

Indoor use only

Operating temperature: 15°C to 30°C (59°F to 86°F)

Humidity: 20% to 80%, noncondensing

Altitude: Operation up to 2400 m (7874 ft.) above mean sea level

Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H

Weight: maximum of 25 kg (60 lbs.)

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively

Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240
V~ ± 10%, 0.8 Amp, 47-63 Hz.

AC inlet fuse: 3 Amp, 250 V~, fast acting

Optics

Reporter laser: 532 nm, nominal output 10-15 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)

Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)

Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm

Classification detector: Avalanche photo diodes with
temperature compensation

Doublet discrimination detector: Avalanche photo diodes with
temperature compensation

Fluidics

Sheath flow rate 90 µL ± 5 µL/second

Cuvette: 200 micron square flow channel

Sample injection rate: 1 µL/second

Sample uptake volume: 20-200 µL

Electronics

Reporter channel detection: A/D resolution 14 bits

Communications interface: USB

Luminex XYP instrument, communications interface: RS 232