Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 28

PN 89-00002-00-072 Rev. C

than one background sample, the fluorescence values of the two
background samples are averaged together and subtracted from the
other samples to obtain a net fluorescence intensity.

Samples Tab

The Samples tab lists tests from your batch and displays the results
for each sample.

Figure 5-20 shows a Sample tab example listing three tests. Notice
that the left pane displays Test IL-4, IL-6 and IL-8 with Test IL-4
selected. The right pane shows the IL-4 sample results. Notice that
the well G3 location test result is “<10” and the associated
Comments column for the third sample indicates a sample out of
range error “Sample High/Low” because this sample has an MFI less
than the lowest standard in the standard curve for this non-
competitive batch.

The first sample is within the standard curve range and thus displays
an unflagged test result.

The system does not flag results as normal or abnormal according to
a defined range. Error flags only note samples that fall below or
above the standard curve.

The Comments column cells are editable and the information is
displayed in reports.