Luminex 100 IS Version 2.3 User Manual

Page 193

Advertising
background image

x

MAP Technology

Luminex 100 IS System Setup

PN 89-00002-00-072 Rev. C

B - 11

3. Place the Luminex SD system near the sheath fluid connection

on the left side of the Luminex 100 analyzer. Make the following
connections; refer to Figure B-11 as needed.

Connect the sheath fluid line (blue fitting) to the Sheath Out
connector on the front of the Luminex SD system.

Connect the air line (green fitting) to the Air In connector on
the front of the Luminex SD system.

If you are using the Luminex SD system waste line, connect
the waste line tubing to the waste connector on the left side
of the Luminex 100 analyzer (orange fitting). Place the other
end of the waste line into a large waste receptacle. Cut off
excess tubing. Ensure the waste receptacle is level with the
Luminex 100 analyzer or no more than three feet below it.

Connect the sheath fluid intake line (white fitting) to the
Sheath In connector on the front of the Luminex SD system.

Figure B-11 Luminex Sheath Delivery Device

1. Power and Fault indicators

5. Sheath Out connector (blue)

2. Prime button

6. Sheath In connector (white)

3. Regulator adjust

7. Reservoir window

4. Air In connector (green)

Note:

Do not place the Luminex

SD on top of the Luminex 100
Analyzer.

4.

5.

6.

1.

2.

3.

7.