Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 48

PN 89-00002-00-072 Rev. C

Figure 5-25. Update CAL Targets Dialog Box

10. Repeat step 9 for the CAL2 lot.

11. In the Maintenance tab, click CAL1, then click OK. The device

status section in the status bar changes from “Running” to
“Standby”.

12. Click CAL2, then click OK. Wait until CAL2 completes.

The Device Status section in the status bar changes from “Running”
to “Standby” and the Diagnostics tab turns red. The System Monitor
on the Diagnostics tab displays the date and time in green if CAL1
and CAL2 are successful.

The Diagnostics tab turns red if all substances (CAL or CON) have
not been run and under the following conditions:

• the first time the software is opened

• the first time a system is calibrated

• a new database is installed

• an old database is restored

• a CAL or CON fails.