Export data, Help, Import calibration – Luminex 100 IS Version 2.3 User Manual

Page 75: Import control, Import template, Insert, Invalidate control, Invalidate standard, Load patient list, Log/linear

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 33

Export Data

Exports data to a .csv file in the batch folder. There is no dialog box
to confirm that the data was exported.

Help

Opens the help file. Use this command if you need instructions on
using a feature in the Luminex IS 2.3 software.

Import Calibration

Opens a dialog box in which you select a calibration file that
contains calibration lot information to import to use for calibrating
the analyzer.

Import Control

Opens a dialog box in which you select a control file that contains
control lot information to import for use when verifying calibration.

Import Template

Opens up the Import Template dialog box, from which you select a
template to import for use in a batch.

Insert

This combination menu/command on the Batch Setup window
allows you to add a user-defined number of patients to the command
list, or skip a defined number of wells on the microtiter plate. The
two options on the menu are Acquire Patient and Skip.

Invalidate Control

Invalidates or removes a control from the data analysis. Luminex
does not recommend invalidating controls.

Invalidate Standard

Removes a standard from the data curve. Use this command if doing
so improves the curve fit. Use caution when doing so because
invalidating a standard will greatly affect the curve fit and
subsequently the sample results.

Load Patient List

Opens the Open Patient List File dialog box, from which you select a
patient list text file to append to a batch.

Log/Linear

Toggles the x axis scale on the histogram or dot plot between
logarithmic and linear modes.

Maximize/Minimize

Toggles between maximum and minimum histogram and dot plot
views.

Next Test

This command is found in the Analysis window. Use this command
to view the next test in the batch.