5. EXPLANATION OF BASIC ITEMS

5.4 Easy Setup

Easy Setup tracks parameters that often need to be
configured or changed.

Click [Easy Setup] at the Navigation tree. It shows the
Easy setup window at the Parameter Set Application as
shown in Figure 5.2.5. The following figure is [Easy
Setup] window in the toggle the tree window mode,
and its contents is described Table 5.4.1.

Figure 5.4.1

Easy Setup

Table 5.4.1 Easy Setup Items

T0503.eps

Block name

Item

Type of Entry

TAG

Descriptor

Message

ASCII code

ASCII code

ASCII code

Transducer
Block

Target Mode

Actual Mode

Measured Value Unit

Alternative

Alternative

Monitor only

Filter Time

Numeric

Analog Input

Target Mode

Alternative

Monitor only

Numeric

Numeric

Numeric

Numeric

Actual Mode

Filter Time Const

PV Scale: Lower Value

PV Scale: Upper Value

Out Scale: Upper Value

Velocity Check

Numeric

Out Scale: Lower Value

Select 1

Alternative

Out Scale: Unit

Local Display

Select 2

Numeric

Out Scale: Decimal Point

Select 3

Monitor only

Alternative

Alternative

Alternative

The [Target Mode] in each function block must be set
from [AUTO] to [O/S] before changing parameters.

When the parameter in each item has been changed,
the DTM Works confirms its update, hence, press the
[Enter] key to fix. If the [ESC] key is pressed before
fixing, the data comes back to the set data without
changing.

Figure 5.4.2

Parameter change

The [Target Mode] must be returned [AUTO] when the
parameter setting work was finished.

IMPORTANT

Do not turn power OFF immediately after param-
eter setting. When parameters are saved to the
EEPROM, the redundant processing is executed
for an improvement of reliability. Should the
power be tumed OFF within 60 seconds after
setting of parameters, changed parameters are
not saved and may return to their original values.