Yokogawa AXFA14G/C User Manual
Page 65

IM 01E20F12-01E
9-7
9. PARAMETER LISTS
T0903-3.EPS
Index
Parameter
Valid Range
Initial Value
Description
Write
Mode
77
MASS_FLOW_DENSITY
Setting of the density for mass-flow rate. This parameter is necessary
in situations where t, kg, g, klb or lb has been selected as the mass
unit in VOLUME_FLOW_UNITS. If a mass unit is selected in
VOLUME_FLOW_UNITS and a value of 0 is set for this parameter,
the setting alarm “57: Density SetErr.” will be displayed. In such a
case, ensure that the density is set correctly.
O/S
0 to 32000
0
78
LIMSW_1_
VALUE_D
(DS-34)
This parameter is used in order to indicate the current input value
selected LIMSW_1_TARGET. If the target alarm occurred the value
is set 1. Refore to LIMSW_1_TARGET.
–
0: not set,
set exclude 0
0
Value
Value
79
LIMSW_1_TARGET
This parameter selects the input channel used to LIMSW_1_
VALUE_D.
O/S
0: Volume Flow
Value
0
80
LIMSW_1_SETPOINT
Sets the threshold of limitswitch 1. If the value of
LIMSW_1_ACT_DIRECTION is HIGH LIMIT, limit switch 1 turns ON
when LIMSW_1_TARGET has gone beyond LIMSW_1_SETPOINT.
If the value of LIMSW_1_ACT_DIRECTION is LO LIMIT, limit switch
1 turns ON when LIMSW_1_TARGET has gone below
LIMSW_1_SETPOINT. The unit set in LIMSW_1_UNIT applies.
O/S
-10
81
LIMSW_1_ACT_
DIRECTION
Selects the direction of the limit switch 1’s actions:
1: LO LIMIT (Low-limit switch)
2: HIGH LIMIT (high-limit switch)
O/S
0: Low Limit
0: Low Limit
1: High Limit
82
LIMSW_1_HYSTERESIS
Sets the hysteresis of limit switch 1 to be applied for resetting the
LIMSW_1_VALUE_D to OFF after LIMSW_1_TARGET went beyond
LIMSW_1_SETPOINT and LIMSW_1_VALUE_D turned
ON (when used as a high-limit switch), or after LIMSW_1_TARGET
went below LIMSW_1_SETPOINT and LIMSW_1_VALUE_D turned
ON (when used as a low-limit switch).
O/S
0.0
0 or larger.
83
LIMSW_1_UNIT
Indicate the unit set in LIMSW_1_TARGET.
–
1349: m
3
/h
84
LIMSW_2_
VALUE_D
(DS-34)
This parameter is used in order to indicate the current input value
selected LIMSW_2_TARGET. If the target alarm occurred the value
is set 1. Refore to LIMSW_2_TARGET.
–
0
0: not set,
set exclude 0
85
LIMSW_2_TARGET
Select the input channel to LIMSW_2_VALUE_D.
O/S
0
0: Volume Flow
Value
86
LIMSW_2_SETPOINT
Sets the threshold of limitswitch 2. If the value of LIMSW_2_ACT_
DIRECTION is HIGH LIMIT, limit switch 2 turns ON when LIMSW_
2_TARGET has gone beyond LIMSW_2_SETPOINT. If the value of
LIMSW_2_ACT_DIRECTION is LO LIMIT, limit switch 2 turns ON
when LIMSW_2_TARGET has gone below LIMSW_2_SETPOINT.
The unit set in LIMSW_2_UNIT applies.
O/S
10
87
LIMSW_2_ACT_
DIRECTION
Selects the direction of the limit switch 2’s actions
0: LO LIMIT (Low-limit switch)
1: HIGH LIMIT (high-limit switch)
O/S
1: High Limit
0: Low Limit
1: High Limit
88
LIMSW_2_HYSTERESIS
Sets the hysteresis of limit switch 2 to be applied for resetting the
LIMSW_2_VALUE_D to OFF after LIMSW_2_TARGET went beyond
LIMSW_2_SETPOINT and LIMSW_2_VALUE_D turned ON
(when used as a high-limit switch), or after LIMSW_2_TARGET
went below LIMSW_2_SETPOINT and LIMS W_2_VALUE_D turned
ON (when used as a low-limit switch).
O/S
0
0 or larger.
89
LIMSW_2_UNIT
Indicate the unit set in LIMSW_2_TARGET.
–
1349: m
3
/h
90
SWITCH_1_
VALUE_D
(DS-34)
Indicate the value of switch 1, which switches ON and OFF
depending on the digital value of the target input parameter selected
in SWITCH_1_TARGET.
–
0
0: not set,
set exclude 0
91
SWITCH_1_TARGET
This parameter selects the input channel used to LIMSW_1_
VALUE_D.
O/S
0: Adheshion
Alarm
0: Adheshion Alarm
1: Adhesion Warning
–
Good
Status
–
Good
Status
Value
–
Good
Status