7. DIAGNOSTIC INFORMATION

7.

DIAGNOSTIC INFORMATION

7.1 DEVICE STATUS

Diagnostic information and failures of the AXF are indicated by using parameter DIAGNOSIS (slot 0, index 29) and
DIAGNOSIS_EXTENSION (slot:0, index 30) in Physical Block. Each diagnostic information is supported when the
corresponding bit in DIAGNOSIS_MASK and DIAGNOSIS_EXTENSION is set. Diagnosis and Diagnosis Extension
are listed in Table 7.1 and Table 7.2.

Table 7.1

Contents of DIAGNOSIS

T0701.EPS

1

0

DIA_HW_ELECTR

Hardware failure of the electronic

10:uP Fault

1

17:Comm uP Fault

19:I/F Module Fault

12:A/D(H) Fault

13:A/D(L) Fault

14:A/D(Z) Fault

100:Comm uP Fault

105:Comm Error1

106:Comm Error2

1

DIA_HW_MECH

Hardware failure mechanics

15:Coil Open

1

2

DIA_TEMP_MOTOR

Motor temperature too high

0

3

DIA_TEMP_ELECTR

Electronic temperature too high

0

4

DIA_MEM_CHKSUM

Memory error

11:EEPROM Fault

1

101:Comm EEPROM Fault

5

DIA_MEASUREMENT

Failure in Measurement

30:Sig Overflow

1

6

DIA_NOT_INIT

Device not initialized (No self calibration)

16:EEPROM Dflt

1

7

DIA_INIT_ERR

Self calibration failed

0

2

0

DIA_ZERRO_ERR

Zero point error (limit position)

82:Auto Zero Wng

1

1

DIA_SUPPLY

1 DIA_SUPPLY Power supply failed

0

(electrical, pneumatic)

2

DIA_CONF_INVAL

Configuration not valid

50:Span > 10m/s

1

51:Span < 0.1m/s

57:Dens Set Err

71:Meas Mod Set

72:Size Set Err

73:Adh Set Err

3

DIA_WARMSTART

New-start-up (warm start up) carried out.

hot start

0

4

DIA_COLDSTART

Re-start-up (cold start up) carried out.

HW,SW,DL restart power fail

0

5

DIA_MAINTENANCE

Maintenance required

31:Empty Pipe

1

33:Adhesion Alm

80:Adhesion Wng

85:Flow Vel Over

6 DIA_CHARACT

Characterization

invalid

0

7

IDENT_NUMBER_Violation

Set to 1 (one), if the Ident_Number of the running

1

cyclic data transfer and the value of Physical Block

IDENT_NUMBER_SELECTOR parameter are different.

3

0-7

Reserved

Reserved for use within the PNO

X

4

0-6

Reserved

Reserved for use within the PNO

X

7

EXTENSION_AVAILABLE

More diagnosis information is available.

1

Octet

Mnemonic

Description

Remarks

Mask

Bit