2 optional specifications, Optional specifications -2 – Yokogawa AXFA14G/C User Manual
Page 57

IM 01E20F12-01E
8-2
8. GENERAL SPECIFICATIONS
Enhanced dual frequency excitation
(Option code HF2): Standard accuracy
±
1 mm/s
Repeatability:
±
0.1% of Rate (V
м 1 m/s (3.3 ft/s))
±
0.05% of Rate
±
0.5 mm/s (V
Ͻ 1 m/s (3.3 ft/s))
8.2
OPTIONAL SPECIFICATIONS
For options other than below, refer to IM 01E20D01-01E and IM 01E20C02-01E
(Optional codes C1, C2, C3, EM, G11 and G13 are unable to select).
<Factory Setting>
As specified in order
In case of different Software Tag (TAG NO) is required from Tag Number
above in the amplifier memory, specify at Software Tag. Default (FT2001)
be set for TAG NO unless otherwise both Tag Number and Software Tag
specified in order.
‘0x7E’ unless otherwise specified in order
Always set as 'No Linearization'
FROWRATE SPAN of flowtube order information be set in PV SCALE.
Lower Range Value be always zero.
Refer to Table below.
'OUT SCALE' always be the same as 'PV SCALE'.
Tag Number (Name Plate and/or stainless steel tag plate)
Software Tag (TAG NO)
Node ADDRESS (Bus Address)
Output Mode (Characterization Type)
Calibration Range (PV SCALE) Lower/Higher Range Value
Calibration Range Unit
Output Scale (OUT SCALE) Lower/Higher Range Value
T05.EPS
MODEL AND SUFFIX CODE
Integral Flowmeter AXF:
AXF
ٗٗٗٗ-Gٗٗٗٗٗ-ٗٗٗٗ-ٗٗٗ/ٗ
Remote Converter AXFA14:
AXFA14
ٗ-Gٗ-ٗٗ/ٗ
(Note1) “G” following the first dash indicates that the output is
digital communication compliant with the PROFIBUS
PA protocol.