<2. Specifications>

2-2

IM 11M10B01-01E

Contact output:

3 outputs

Error contact, O

2

concentration high/low alarm contact, range marker output

Multi selector (optional)

: Contact output for switching sample gas flow, measument flow information contact output.

Note : For details , see external dimensions.

Contact output specifications

General

: Relay

Nominal contact capacity:

2 A 30V DC, 2 A 240 V AC (120 V AC for 100 V power supply) for resistive load

Maximum power: 60 W, 480 VA
Maximum voltage: 30 V DC, 264 V AC (132 V AC for 100 V power supply)
Maximum current: 2 A DC/AC

Contact input

: Voltage-free contact, 1 point

: Remote switching for sample gas suction pump ON/OFF

Self-diagnostics

Error (failure) : Sensor unit error, heater temperature error, temperature sensor disconnection,

device temperature error, CPU error, fan stop.

Alarm (warning) : Heater unstable, sensor defect, electromotive force abnormal, asymmetry voltage

error, calibration error, sensor resistance error, O

2

concentration upper/lower, over range.

Serial communication

Comm. signal : RS-232 , one way

Baud rate

: 38,400 bps

Data (ASCII)

: O

2

concentration, unit, alarm/error

Calibration methods:

(1) 3 point:

10 ppm, 1000 ppm, Air

(2) 2 point:

zero and span calibration may be set freely

(3) 1 point:
(4) Air calibration

Warm-up time

: Within 20 min

Power supply:

Power supply : 100 - 120 V AC/200 - 240 V AC, 50/60 Hz
Acceptable range: 100 to 120 V AC ±10%, 50/60 Hz

200 to 240 V AC ±10%, 50/60 Hz

Power consumption

: 100 to 120 V AC, 200 VA max.
200 to 240 V AC, 400 VA max.

Dimensions

: 213 (W) x 132 (H) x approx. 375 (D) mm

Weight : Approx. 5 kg

Finish : Polyester coating