<6. Operation and Parameters>

6-9

IM 11M10B01-01E

and press the [ENT] key.
Check the current calibration coefficient. These coefficients are updated for each calibration. Figure

6.11 shows an example of checking the coefficient.

CoEF

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT] 1. 011

dS1

PPM

[ENT] - 0. 23

dZ1

PPM

[ENT]

▼↓↑▲

1. 013

dS4

1. 35

dZ4

[ENT]

PPM

[ENT]

PPM

MODE

%

%

%

%

To “rEV”

Hold down

for 2 seconds.

Return to “CoEF.”

Figure 6.11

Checking Calibration Coefficient “CoEF”

Figure 6.12 shows an example of displaying the cell resistance value.

CEL.r

MODE

CAL

▼↓↑▲

▼↓↑▲

[ENT]

56. 1

CEL.r

PPM

[ENT]

▼↓↑▲

%

To “rEV”

Hold down

for 2 seconds.

Return to “CEL.r.”

Figure 6.12

Displaying Cell Resistance Value “CEL.r”

6.1.12 Displaying Software Revision “rEV”

This item is for checking the software revision of the OX400 and is not a setting.
Operation: Hold down the [MODE] key for 2 seconds, select “rEV” with the [▲], [▼] keys, and

press the [ENT] key.

The current software revision is displayed. Figure 6.13 shows a display example.

6.1.11 Displaying Cell Resistance Value “CEL.r”

This item is for checking the sensor state and is not a setting.
Operation: Hold down the [MODE] key for 2 seconds, select “CEL.r” with the [▲], [▼] keys,

and press the [ENT] key. The sensor cell resistance value is displayed. The unit is Ω. If the

value becomes larger than 1050 Ω, “oVER” appears.

NOTE

When the sensor is replaced, the cell resistance value is not displayed. It is displayed when calibration

of something other than air is performed.