<2. Specifications>

2-11

IM 11M10B01-01E

2.7 Wiring Diagram

MS1

MS2

MS3

MSI1

MSI2

MSICOM

MSCOM

1st flow path

2nd flow path

3rd flow path

OX400 Rear Terminal (M3 screw)

The following terminals are added for the multi-selector “/MS” option.

The customer needs to supply a switching device and carry out the necessary wiring.

Contact output
for switching of
measurement flow
path

Contact output
for measurement
flow path data

Common

Common

Common

PUMP

OFF

mA

G

V

RC2
RC3
RCCOM

RC1

FAIL

DO

Pump ON/OFF

Contact input

(available when suffix code of

with built-in pump “-P” specified)

Secondry output
(0-1/0-5/0-10 V DC)

Power cord (two-pole plug with earthing contact)

Primary output
(4 to 20 mA DC)

+

-

+

-

Alarm

(Oxygen concentration high/low)

Contact output

Error (Fail)

Contact output

Range marker

Contact output

(Note 2) (Note 3)

(Note 1)

Note 1: Use the earthing contact of power cord to ground to earth.

Use the supplied power cord only.

Note 2: Ground the measurement output signal line shield on the receiving side.

The G-terminal is connected to a ground pin. Use this if the line shield cannot be grounded on the receiving side.

Be very careful not to ground the line at two points.

Note 3: The signal output and contact input cable must be no longer than 30 m for CE marking. RS232 cable must be no longer than 3 m.