3 mounting activated carbon filter, Mounting activated carbon filter -4 – Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 27

Advertising
background image

<3. Installation>

3-4

IM 11M10B01-01E

3.2.3 Mounting Activated Carbon Filter

Mount the filter on a panel or wall. To fix the filter with M5 screws. When replace a filter pack, 200 mm

or more of maintenance space is necessary. If necessary, the piping connection should be removable.

(Rc1/4 or 1/4NPT)

(Rc1/4 or 1/4NPT)

110

K9643KH: Rc1/4

K9643KJ: 1/4NPT

± 0.5

± 0.5

Unit: mm

Sample gas outlet

2-Ф6.5 holes

2-M5 screws

Mark “A”

In case of

K9643KJ

Activated carbon filter of

option code “/A”

Sample gas inlet

Fixing holes

more than 200

Maintenance space

The filter holder including

piping connection can be

moved upward to replace

a filter element.

more than

200

Maintenance space

The filter holder including

piping connection can be

moved downward to

replace a filter element.

110

150

130

47

29.5

25

15

10

(157)

35

35

60

47

Figure 3.3

Mounting Activated Carbon Filter

NOTE

The activated carbon filter used in this equipment have limited life, and periodic maintenance or

replacement will be required. Install this equipment in an accessible location to facilitate maintenance.
Maintenance conditions vary depending on the gas conditions. Replacement intervals for the filter and

activated carbon should be determined in accordance with the operating conditions.
• Activated carbon filter life is about 50 hours. (if measured gas contains isopropyl alcohol of

concentration 500 ppm, and flow is 500 ml/min.)

• In case of using the built-in pump or with the aspirator, the activated carbon filter life is about 25

hours.