<4. Piping and Wiring>

4-2

IM 11M10B01-01E

!

INLET

OUTLET

N200

C

US

R

172608

SERIAL

(RS-232)

FUSE

250V T3.15A

!

RCCOM

RC3

RC2

RC1

FAIL

DO

PUMP

OFF

G

V

-

+

mA

-

+

MS1

MS2

MS3

MS

COM

MSI

COM

MSI2

MSI1

G

+

-

-

For piping, use a metal pipe.

Screw in the pipe while holding the hexagonal part with

a wrench or the like. Not doing so and using a strong force

when screwing in the pipe may cause the thread on the OX400

to rotate, resulting in damage to the internal pipe.

Gas inlet

Gas outlet

(The connection is

Rc1/4 or 1/4NPT)

CAUTION

Figure 4.2

Piping for Sample Gas

4.2

Wirings

4.2.1 Power Supply Wiring

CAUTION

For the power supply, be sure to use the rated voltage shown on the rear panel. Connection to a

power supply with a different voltage may cause damage to the OX400. Power cord connector with

a protective ground pin is used for power supply wiring for the OX400. Be sure to securely insert the

connector of supplied power supply cord to the OX400, and connect the plug to a 3P socket with a

protective ground pin.

WARNING

Before connecting wires, be sure to remove the power plug from the socket and check that the power

is turned off in order to prevent electric shock. Furthermore, after finishing the wiring, be sure to secure

the special terminal cover with screws.

• Do not use the supplied power cord with another device.