3.2.2 Installing Panel Mount Type

(1) Attach a panel mount frame to the side of the OX400, insert it into the panel, and securely screw

it to the panel.

(2) Provide a distance of 100 mm or more behind the OX400 in order to not block the outlet of the

cooling fan on the rear panel.

(3) The air inlet of the cooling fan is located on the bottom panel of the OX400. Be sure to provide

a distance of 25 mm or more from the bottom panel in order to not block the bottom panel of the

OX400.

135

+2 0

101.6

±0.3

240 ±1

225 ±1

19.3

330

20

10 to 20

6

6

132

101.6

260

238

Cooling fan

Unit: mm

Do not block
the air inlet.

Air is discharged

from the cooling

fan

Distance of 100 or more

Seal

Frame

Frame

Screw

4xM5 screw

Panel Cutout

Mount

3. Make sure the panel is at least 2 mm thick.

2. Maintain at least 100 mm of free space around the measuring instrument in order to ensure

adequate ventilation.

1. Make sure the bottom supports do not block the ventilation outlet on the bottom panel of

the measuring instrument.

Notes on mounting

Distance of 25 or more

Figure 3.2

Notes on Mounting Panel Mount Type