2 setting secondary output “out2, 3 setting hold function “hold, Setting secondary output “out2” -4 – Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 41: Setting hold function “hold” -4

Advertising
background image

<6. Operation and Parameters>

6-4

IM 11M10B01-01E

6.1.2 Setting Secondary Output “oUt2”

Operation: Hold down the [MODE] key for 2 seconds, select “oUt2” with the [▲], [▼] keys, and

press the [ENT] key.
With respect to the secondary output, select one of the three types of voltage output: 0-1 V is “1,” 0-5

V is “5,” and 0-10 V DC is “10.” Figure 6.3 shows an example of setting 0-1 V DC “1.”

CAL

rnG

oUt2

HoLd

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

1

oUt2

PPM

5

oUt2

10

oUt2

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

1

oUt2

PPM

[ENT]

MODE

(0-10 V D C )

(0-5 V D C )

%

%

To “rEV”

Hold down

for 2 seconds.

To “10”

0-1 V DC is confirmed.

Return to “oUt2.”

To “1”

Figure 6.3

Setting Secondary Output “oUt2”

6.1.3 Setting HOLD Function “HoLd”

Operation: Press the [MODE] key for 2 seconds, select HoLd with the [▲], [▼] keys, and press

the [ENT] key.
Set the output state in the maintenance mode. Select either “non” or “PrEV.” The former option means

no hold, and the latter means hold the previous value. The following is an example of no hold.

HoLd

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT]

HoLd

PPM

HoLd

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

HoLd

PPM

[ENT]

MODE

non

PrEV

non

To “rEV”

Hold down

for 2 seconds.

To “PrEV”

No hold is confirmed.

Return to “HoLd.”

(Hold the previous value.)

To “non”

%

%

Figure 6.4

Setting HOLD “HoLd”