5. Names and Functions>

5-1

IM 11M10B01-01E

5. Names and Functions

5.1 Front Panel

1

2

3

4

5

6

8

9

10

11

12

7

Figure 5.1 Description of Front Panel

1

Main Display

The main display digitally displays the oxygen concentration and set value in four digits with a

decimal point. The display flashes at one-second intervals while waiting for key input.

2

Oxygen Concentration Unit LED

The unit of displayed oxygen concentration is displayed by LED. The LED is not on when a non-

oxygen concentration value is displayed. The unit you selected flashes until you confirm it.

3

MODE LED

The LED turns on when you push the MODE switch to enable the maintenance mode. The LED

is off in the measurement mode.

4

ERR/ALM LED

The event of an error causes the LED to turn on. The event of an alarm causes the LED to flash.

If an error and alarm occur at the same time, the error is given priority and the LED turns on. The

LED is off during normal operation.

5

Sub-Display

The sub-display displays the description of the currently displayed item or menu item in four

digits with a decimal point. If an error or alarm occurs, the error or alarm code corresponding to

the content is displayed. “HEAt” is displayed during the warm-up period.

6

Flowmeter

The flowmeter uses a floater to indicate the flow rate of gas flowing into the sensor. Always set

the value to 200 ±25 ml/min (a floater should be between upper and lower bar of the 200 ml/min

bar on flowmeter).

7

MODE Key

Hold down the key for 2 seconds to switch from the measurement mode to the maintenance

mode (or from the maintenance mode to the measurement mode).