<6. Operation and Parameters>

6-6

IM 11M10B01-01E

6.1.5 Setting Alarms (Oxygen Concentration High/Low Alarms) “ALM”

Operation: Hold down the [MODE] key for 2 seconds, select “ALM” with the [▲], [▼] keys, and

press the [ENT] key.
High and low limit alarms can be set to the measured oxygen concentration values. With respect to

the alarm setting, the following four options are available: No alarm “oFF,” high/low limits “ALL,” high

limit alarm “AL.HI,” and low limit alarm “AL.Lo.” If the high and low limit values are reversed or the set

value is larger than 100%, an input error occurs and the setting is not accepted. Furthermore, the DO

contact output closes in the event of an oxygen concentration alarm (ALM7). In other words, an alarm

contact is activated.
Figure 6.6 shows a setting example. With respect to the input method of a numerical value, decimal

point, and unit, see Section 6.6.

CAL

ALM

▼↓↑▲

▼↓↑▲

[ENT]

oFF

ALM

[ENT]

[ENT]

▼↓↑▲

oFF

ALM

▼↓↑▲

▼↓↑▲

▼ ▲

▼ ▲

100. 0

AL.HI

PPM [ENT]

1

100 00.11

AL.Lo

PPM

[ENT]

[ENT]

0

0.

000

AL.Lo

PPM [ENT]

0. 000

AL.Lo

PPM[ENT]

ALL

ALM

1

1

00. 0

AL.HI

PPM

HI

ALM

▼ ▲

Lo

ALM

[ENT]

0

0.

000

AL.Lo

PPM [ENT]

0. 000

AL.Lo

PPM [ENT]

[ENT]

1

1

00

.

0

AL.HI

PPM [ENT]

100. 0

AL.HI

PPM [ENT]

MODE

%

%

%

%

%

%

%

%

%

To “rEV”

Hold down for 2 seconds.

To “Lo”

No concentration alarm is confirmed.

Return to “ALM.”

Return to “ALM.”

Return to “ALM.”

Return to “ALM.”

Enter the concentration

alarm low limit value.

Enter the numerical value,

decimal point, and unit.

Enter the numerical value,

decimal point, and unit.

Enter the numerical value,

decimal point, and unit.

Enter the numerical value,

decimal point, and unit.

Low limit value

is confirmed.

Enter the concentration

alarm high limit value.

High limit value is confirmed.

The main display flashes

in the event of an input error.

Examples of input error

High and low limit values are reversed.

Input value is larger than 100%.

Pressing the [MODE] key in the event of an input

error returns to the previous operation.

Enter the concentration alarm high limit value.

High limit value is confirmed.

Enter the concentration alarm low limit value.

Low limit value is confirmed.

To “oFF”

Figure 6.6

Setting Alarm (Oxygen Concentration Alarm) “ALM”