8 setting output smoothing “smoo, 9 setting multi-selector “mls, 10 checking calibration coefficient “coef – Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 45: Smoo

Advertising
background image

<6. Operation and Parameters>

6-8

IM 11M10B01-01E

6.1.8 Setting Output Smoothing “SMoo”

If the oxygen concentration of sample gas rapidly changes, and the measured value is used for

control, harmful results may occur such as frequent on and off switching.
In such a case, signal changes can be smoothed by giving an appropriate time constant and

performing calculation. Smoothing factor from 0 to 60 seconds can be set. When the output smoothing

is set, the smoothed value is an instantaneous value, which is displayed and output in analog form.

The smoothed value is also used as the instantaneous value for checking the concentration alarm and

for RS232 communication. The output smoothing is “non“ 0 second by default.

Operation:

Hold down the [MODE] key for 2 seconds, select “SMoo” with the [▲], [▼] keys,

and press the [ENT] key.

Figure 6.9 shows an example of setting the smoothing factor to 30 seconds.

SMoo

SMoo

[ENT]

[ENT]

SMoo

00

30

Entering 30 sec. of smoothing factor.

Enter “30” and

press the

[ENT] key.

Return to “SMoo.”

MODE

Hold down

for 2 seconds.

Figure 6.9

Setting Smoothing Factor “SMoo”

6.1.9 Setting Multi-Selector “MLS”

This item is an option, so you do not need to set it unless you specified the option code “/MS.”
Operation: Hold down the [MODE] key for 2 seconds, select “MLS” with the [▲], [▼] keys, and

press the [ENT] key.
Set the relay contact to switch measurement flow for path. Up to three sampling flows can be

selected. Even though this setting is performed for an OX400 for which the option code “/MS” was not

specified, nothing will happen. Figure 6.10 shows a setting example (set measurement flow path 1 to

3). With respect to the usage example, see Section 6.5.

MLS

▼↓↑▲

MODE

CAL

▼↓↑▲

▼↓↑▲

[ENT]

3

MLS

PPM

0

MLS

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

0

MLS

PPM

[ENT]

▼↓↑▲

▼↓↑▲

%

%

To “rEV”

Hold down

for 2 seconds.

To “0”

0 is confirmed.

Return to “MLS.”

To “3”

Figure 6.10

Setting Multi-Selector “MLS”

6.1.10 Checking Calibration Coefficient “CoEF”

This item is for checking the sensor state and is not a setting.
Operation: Hold down the [MODE] key for 2 seconds, select “CoEF” with the [▲], [▼] keys,

Advertising