<2. Specifications>

2-3

IM 11M10B01-01E

Line connection

Gas inlet : Rc1/4 or 1/4NPT female
Gas outlet : Rc1/4 or 1/4NPT female

Electrical connection

External output terminals: M3 screw
Contact input/output terminals: M3 screw
Serial communication: D-sub 9 pin connector
Ground: within power cord connector

Environment and operational conditions

Installation conditions: Indoors, panel or wall mounting, non explosion area

Ambient temperature: 0 to 40°C, non-condensing

Ambient humidity: 5 to 85% RH

Storage temperature: -5 to 50°C

Conformance to

Safety and EMC standards

US

C

R

Safety : EN 61010-1

CAN/CSA-C22.2 No. 61010-1-04, UL Std. No. 61010-1

EMC: EN 61326-1 Class A, Table 2 (for use in industrial locations)

EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
(Note1) This instrument is a Class A product, and it is designed for use in the industrial environment.

Please use this instrument in the industrial environment only.

(Note1) The current or voltage signal output cable length and contact input cable length must be no

longer than 30 m for CE marking. RS232 connection cable length must be no longer than 3 m.

Installation altitude:

2000 m or less

Category based on IEC 61010:

II (Note)

Pollution degree based on IEC 61010: 2 (Note)

Note: Installation category, called over-voltage category, specifies impulse withstand voltage.

Category II is for electrical equipment.

Pollution degree indicates the degree of existence of solid, liquid, gas or other

inclusions which may reduce dielectric strength. Pollution degree 2 is the normal indoor

environment.

2.2 Characteristics

Repeatability

: ±1% FS (Hereafter, either 10, 100, 1000 ppm, 1%, 10%, or 100% O

2

is FS)

Linearity

: ±2% FS
±3% FS (0-100 ppm or less)