<6. Operation and Parameters>

6-5

IM 11M10B01-01E

6.1.4 Setting Burnout Function “nAMU”

Operation: Hold down the [MODE] key for 2 seconds, select “nAMU” with the [▲], [▼] keys,

and press the [ENT] key.
Set the burnout function for the primary current output of 4-20 mA DC (compliant with NAMUR). Select

one of the options: No burnout function is “non,” burn-up is “Er.HI,” and burn-down is “Er.Lo.”
Figure 6.5 shows an example of setting burn-up.
NOTE: Burnout is a function compliant with NAMUR that allows increasing/decreasing the current

output to the high/low limits in the event of a failure. Burn-up allows increasing the output to the

high limit of 21.0 mA and burn-down allows decreasing the output to the low limit of 3.6 mA.

The burnout function works in the event of an error (when the FAIL contact output closes).

nAMU

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT]

nAMU

PPM

nAMU

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

E

Er. H

HI

nAMU

PPM

[ENT]

nAMU

▼↓↑▲

▼↓↑▲

ALM

▼↓↑▲

MODE

To “Er.Lo” burn-down

Burn-up is confirmed.

Return to “nAMU.”

To “non”

n

noonn

E

Err. HHI

E

Err. LLoo

%

To “rEV”

Hold down

for 2 seconds.

%

Figure 6.5

Setting Burnout “nAMU”