Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 57

Advertising
background image

<

7. Inspection and Maintenance>

7-4

IM 11M10B01-01E

5) Remove the lead wire connector of the sensor from the PCB board.
6) Loosen the joint ring

1

on the sensor outlet side and remove it towards the heater side.

7) Pull out the joint

2

while holding the sensor.

8) Rotate the inlet-side joint nut

3

counter clockwise seen from the heater side and remove it.

O-ring

Sensor

Lead wire connector

Lead wire connector

Heater

Heater

1

1

2

2

3

3

9) Pull out the holding plate

4

, remove the ring

1

, and pull the sensor out of the heater

straightforward.

O-ring

Lead wire connector

Heater

1

2

3

4

Figure 7.2

Removing Sensor

Advertising