6 piping diagram, Piping diagram -10 – Yokogawa Low Concentration (ppm) Zirconia Oxygen Analyzer OX400 User Manual

Page 21

Advertising
background image

<2. Specifications>

2-10

IM 11M10B01-01E

2.6 Piping Diagram

P

P

Flow meter

Suction pump

Heater

Oxygen

sensor

IN

OUT

IN

OUT

Oxygen

sensor

With aspirator (Sampling method [-A])

Built-in pump (Sampling method [-P])

Filter1

Filter2

Restriction

Filter1

Filter2

Aspirator

Sample gas

Sample gas

Pipe connection
Rc1/4 or
1/4NPT

Air or N

2

supply gas

inlet Φ6 connector

Φ8 connector

(Operable from

the front panel)

Note: If no suction device [-N] is specified, the aspirator is removed from this diagram.

Flow meter

Heater

Restriction

(Operable from

the front panel)

Throttle

valve

Throttle

valve