<6. Operation and Parameters>

6-1

IM 11M10B01-01E

6. Operation and Parameters

The OX400 has two modes, “Measurement Mode” and “Maintenance Mode.” Oxygen concentration

is displayed in the measurement mode. The setting of operation parameters and calibration operation

are performed in the maintenance mode. Furthermore, the MODE LED on the front panel is on in the

maintenance mode. To enter the maintenance mode, or to return to the measurement mode from the

maintenance mode, hold down the [

MODE] key for two seconds. The parameters shown in Figure 6.1

are available in the maintenance mode. Parameters can be displayed by scrolling with the [▼], [▲]

keys.

MODE

▼↓↑▲

CAL

Hold down

for 2 seconds

Hold down

for 2 seconds

CAL

▼↓↑▲

[ENT]

rnG

oUt2

HoLd

nAMU

ALM

SEt.C

CoEF

rEV

SEnS

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

MODE

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

CEL.r

▼↓↑▲

[ENT]

Performs calibration.

(3-point / 2-point / 1-point / Air)

Destination

(reference section)

Maintenance mode

“0-1 V DC”

To “rEV”

To measurement

mode

Description

Default

To Calibration “CAL”

(Section 6.2)

To Displaying Cell

Resistance Value

(Section 6.1.11)
To Setting

Output Range

(Section 6.1.1)

To Setting Secondary

Output (Section 6.1.2)

To Setting Hold

(Section 6.1.3)

To Setting Burnout

(Section 6.1.4)

To Setting Alarms

(Section 6.1.5)

To Setting Calibration

Gas Concentration

(Section 6.1.6)

To Displaying

Calibration Coefficient

(Section 6.1.10)
To Setting Sensor

Constant

(Section 6.1.7)
To Displaying Software

Revision “rEV”

(Section 6.1.12)

For maintenance and diagnosis.

For maintenance and diagnosis.

Sets the

measurement range.

(Auto / Fixed / Partial)
Selects the voltage output.

(0-1 / 0-5 / 0-10 V DC)
Specifies the output hold

during maintenance.

(None/Previous value)

Specifies the burnout function

in the event of a failure

(None / Burn-up / Burn-down)
Sets the oxygen

concentration alarm (DO).

(None / HL&LL / Only HL / Only LL)
Configures the setting at the

time of calibration. (Note 1)

(See 6.2 Calibration “CAL”

when configuring the setting

while performing calibration)

The setting must be configured

when replacing sensors.

(dZ1 / dS1 / dZ4 / dS4 / Con4)

Displays the revision of the

software for maintenance.

“All”

(3-point)

“Auto”

“None”

“None”

“None”

Note 1: When “Air” is selected in Calibration “Cal,” the “SEt.C” menu is not displayed.

Note 2: Some settings are omitted in the above flow chart.

For setting output smoothing function of “SMoo” menu, see Section 6.1.8.

When switching of measurement flow path using the Multi-selector,

set the measurement flow path (1 to 3) in “MLS.” For the operation of the “MLS” menu,

see Section 6.1.9.

Figure 6.1

List of Parameters in Maintenance Mode