<6. Operation and Parameters>

6-7

IM 11M10B01-01E

6.1.6 Setting Calibration Gas Concentration “SEt.C”

Operation: Hold down the [MODE] key for 2 seconds, select “SEt.C” with the [▲], [▼] keys,

and press the [ENT] key.
Set the O

2

concentration of gas used for calibration. Alternatively, this setting can be performed

while performing calibration “CAL.” The gas concentration that can be set is restricted depending on

which calibration is performed (which depends on the specification in calibration “CAL”). The oxygen

concentration of air is fixed to 20.6% O

2

, so it cannot be set. Figure 6.7 shows a setting example in 3-

point calibration “ALL.” With respect to the input method of a numerical value, decimal point, and unit,

see Section 6.6.

SEt.C

▼↓↑▲

MODE

CAL

▼↓↑▲

▼↓↑▲

10.00

PPM

10.00

PPM

1000

1000

PPM

PPM

%

%

%

%

10. 00

1000

10. 00

1000

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

To “rEV”

Hold down

for 2 seconds.

Return to “SEt.C.”

10 ppm gas concentration

is confirmed.

1000 ppm gas concentration

is confirmed.

Figure 6.7

Setting Calibration Gas Concentration “SEt.C”

6.1.7 Setting Sensor Constant “SEnS”

Operation: Hold down the [MODE] key for 2 seconds, select “SEnS” with the [▲], [▼] keys,

and press the [ENT] key.
Be sure to enter the sensor constant when replacing sensors. With respect to the sensor constant,

the following five parameters are available: dZ1, dS1, dZ4, dS4, and Con4, which are on the constant

tag of the sensor. Be sure to enter the right parameter. Figure 6.8 shows a setting example. These

constants are specific to the sensor and remain the same for all operations. With respect to the input

method of a numerical value, decimal point, and unit, see Section 6.6.

dZ1

00Ab

dZ1

dS1

AE0E

dS1

Con4

PPM

1021

Con4

PPM

dZ4

FF79

dZ4

dS4

AC75

dS4

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

%

%

SEnS

0

0

A

1

F

A

000

0AC

E0F

F7C

C77

038

Entering password

Enter “0330” and press the [ENT] key.

Enter the zero sensor constant 1

(in hexadecimal). Input is confirmed.

Input is confirmed.

Input is confirmed.

Input is confirmed.

Input is confirmed.

Enter the span sensor constant 1

(in hexadecimal).

Enter the zero sensor constant 4

(in hexadecimal).

Enter the span sensor constant 4

(in hexadecimal).

Enter the gas concentration of Con4 (1000 ppm).

Enter the numerical

value only.

Return to “SEnS.”

*The entered sensor constants are updated altogether when the input of Con4 is confirmed.

If the input is interrupted halfway, it returns to the sensor constant before the input.

SEnS

Figure 6.8

Setting Sensor Constants “SEnS”